CON 600

Product Description

pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample

The selection of a suitable electrode depends on the types of application

Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable

Single Junction

Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers

Double Junction

Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes

Sealed(gel-filled)

No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required

Refillable

Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low

Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses

pH Electrodes (General Glass) & (General Plastic)

 

Description

  • Study waterproof design suitable for outdoor use
  • Wide measuring range, appropriate range selection enabled by auto ranging capabilities of the meter
  • High accuracy with ±1 % full scale accuracy at 3 decimal resolutions
  • Backlit LCD screen with multi data display
  • Cal Due alarm for periodic calibration updates
  • Meter logs reading automatically within seconds of measurements
  • It allows you to automatically log up to 500 data sets in GLP-compliant format with time and date stamping
  • RS 232 through LED, USB / IrDA Interface Adapter
  • Data transfer is easy with the inbuilt infrared IRDA and RS232 communication technology

Product Description

Product Specification Description
Conductivity Range 0 to 2.000 μS/cm,2.000 μS/cm to 300.0 μS/cm,
300.0 μS/cm to 4.000 mS/cm,4.000 mS/cm to 40.00 mS/cm,40.00 mS/cm to 200.0 mS/cm
Resolution / Accuracy 0.01 μS/cm, 0.1 μS/cm, 0.001 mS/cm, 0.01 mS/cm, 0.1 mS/cm (0.05% of full scale) ; ±1 % full scale of each range
TDS Range 0 to 2.000 ppm,2.000 ppm to 300.0 ppm,300.0 ppm to 4.000 ppt,4.000 ppt to 40.00 ppt,40.00 ppt to 200.0 ppt Factor : 0.40 to 1.00
Resolution / Accuracy 0.01 ppm, 0.1 ppm, 0.001 ppt, 0.01 ppt, 0.1 ppt (0.05% of full scale) ; ±1 % full scale of each range
Temperature Range -10.0 to 110.0 °C/14.0 to 230.0 °F (Meter)
Resolution/Accuracy   0.1 °C / 0.1 °F ; ± 0.5 °C /± 0.9 °F
Temperature Compensation ATC / MTC (0 to 100 °C)
Temperature Coefficient Linear
Temperature Normalization 15 or 30 °C
Cell Constant 0.100 to 10.000
No. of cal. Points 4(1 per range) Auto; 5(1 per range) Manual
Cal. Due Alarm Yes
Auto-Off 2 to 30 Minutes after last key pressed (selectable)
Automatic Endpoint Yes
Data logging 500 Data sets (Non-Volatile memory)
GLP Yes
Communication USB / IrDA Interface Adapter  (for 600 Series Meter to PC Communication) to be ordered separately
LCD Display Back-lit dot matrix ;7.1 cm (H) x 5.4 cm (W)
Input Conductivity 8 Pin connector
Electrode Included 4-cell cond. electrode ECCONSEN9203J,
K = 0.530, 3 m cable length
Ingress Protection IP67
Power Included 4 x 1.5 V “AA” Alkaline batteries or
9 VDC adaptor, 500 mA
Carrying case Yes
Dimensions
(L x W x H)
Meter 18.3 x 9.5 x 5.7 cm
Boxed 23.5  x 16 x 15.5 cm
Weight Meter 460 g
Boxed 800 g

Ordering Information

Order No. Item
ECCONWP60043K Waterproof CyberScan CON 600 Conductivity/TDS Meter with 3m cable  4-cell electrode ECCONSEN9203J K=0.530, DAS software,  power adapter & Conductivity/TDS carrying kit set
01X447602 USB / IrDA Interface Adapter  (for 600 Series Meter to PC Communication)

Conductivity Calibration Solutions

Order No. Item
ECCON100BT 100 μS KCL, 480 ml
ECCON500BT 500 μS KCL, 480 ml
ECCON1413BT 1413 μS KCL, 480 ml
ECCON2764BT 2764 μS KCL, 480 ml
ECCON5000BT 5.0 mS KCL, 480 ml
ECCON1288BT 12.88 mS KCL, 480 ml
ECCON1118BT 111.8 mS KCL, 480 ml

TDS Calibration Solutions

Order No. Item
EC44250BT 50 ppm, 480 ml
EC442300BT 300 ppm, 480 ml
EC4421000BT 1000 ppm, 480 ml
EC4423000BT 3000 ppm, 480 ml