DO 6+

Product Description

pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample

The selection of a suitable electrode depends on the types of application

Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable

Single Junction

Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers

Double Junction

Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes

Sealed(gel-filled)

No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required

Refillable

Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low

Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses

pH Electrodes (General Glass) & (General Plastic)

 

SKU: f001-1-1-2-1-3-1-1-1-2-1-2-1-1-1-1-1-1-2-1-2-2-2 Categories: ,

Description

  • Microprocessor based dissolved oxygen meter. Reads in mg/l, ppm and % Saturation
  • Barometric Pressure and Salinity Compensation
  • Calibration can be performed at 100% and or 0% solution
  • Rugged rubber boot for additional protection
  • Galvanic DO probe allows instantaneous measurements since no time for polarisation is required
  • Easy troubleshooting with comprehensive self diagnostic message

Product Description

Product Specification Description
Waterproof No
Dissolved Oxygen Range 0.00 to 19.99 mg/l or ppm
Resolution/ Accuracy 0.01 mg/l or ppm; ± 1.5% of full scale
% Saturation of Oxygen 0.0  – 199.9%
Resolution/ Accuracy 0.1 % ; ± 1.5% of full scale
Temperature Range 0 to 50.0 0C
Resolution/ Accuracy 0.1 0C ; ± 0.5 0C
Barometric Pressure Correction 500 to 1499 mm Hg or 66.6 to 199.9 kPa
Resolution 1 mm Hg or 0.1 kPa
Method Automatic correction after manual input
Salinity Correction 0.0 to 50.0 ppt
Resolution 0.1 ppt
Method Automatic correction after manual input
Probe Included, galvanic probe with 3 meter cable (ECDO6HANDY3M)
Temp. Compensation Automatic /Manual (0 to 50.0 0C)
Operating Range 0 to 50 0C
Display Custom single 4 digit LCD
Power Requirement 4 x 1.5V ‘AAA’ batteries, battery life > 100 hours. Mains operation not possible
Carrying Case Yes
Dimensions
(L x W x H)
Meter 15.7 x 8.5 x 4.2 cm
Boxed 24 x 17 x 8 cm
Weight Meter 255 g
Boxed 555 g

Ordering Information

Order No. Item
ECDO602PLUSK Galvanic Dissolved Oxygen Meter with 3m cable electrode , 2 assembled membrane cap housing, 1 refilling electrolyte (10ml) & carrying kit set

Zero Oxygen Solution

Order No. Item
ECDOZEROSOLN Zero Oxygen Solution, 0% DO, 480 ml