PC 450

Product Description

pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample

The selection of a suitable electrode depends on the types of application

Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable

Single Junction

Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers

Double Junction

Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes

Sealed(gel-filled)

No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required

Refillable

Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low

Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses

pH Electrodes (General Glass) & (General Plastic)

 

SKU: f001-1-1-2-1-3-1-1-1-2-1-2-1-1-1-1-1-1-2-1-2-2-2-1-1-1-1 Categories: ,

Description

  • Durable waterproof and dustproof design with IP67 rating
  • Switch between pH, mV, conductivity, and TDS with a press of a button
  • Large, backlit custom display – Lets you view measurement with temperature
  • Grip-Clip™ – Quickly and easily secures separate pH and conductivity electrodes simultaneously to your sample container or beaker
  • Calibrate up to 5 points automatically with USA or NIST buffer groups (up to 6 points with DIN). New manual option allows calibration with up to 5 custom buffer standards you
  • Auto buffer recognition – Automatically identifies the correct pH buffer for rapid calibration
  • Calibration due alarm – Set a visible reminder when it is time to recalibrate the meter
  • Auto-range conductivity from 0 to 199.9 mS in five ranges – With up to five calibration points
  • Selectable cell constant – 0.10, 1.00 or 10.0
  • Adjustable temperature coefficient – From 0.00 to 10.00 % per ºC
  • Conductivity-to-TDS factor – Adjustable from 0.40 to 1.00
  • Selectable 0C/ 0F – Choose units to best fit your application
  • Ready indicator, manual or automatic measurement hold feature and optional automatic shutoff function
  • Store up to 500 data sets; easy USB cable, RS232 output with optional data cable
  • Built-in real-time clock stamps stored data and calibration data with time and date – Meets standards for GLP (Good Laboratory Practices)
  • 500 hours battery life or use optional universal power adapter

Product Description

Product Specification Description
pH Range -2.00 to 16.00 pH
Resolution / Accuracy 0.01 pH / ± 0.01 pH
pH Calibration Points Up to 6 points (USA ,NIST ,DIN CUSTOM)
mV Range ± 2000 mV
Resolution / Accuracy 0.1 mV ± 999.9 mV; 1 mV beyond / ±0.2 mV + 2 LSD or ± 0.5 % of reading; whichever is greater
mV Calibration Points ± 200 mV
Conductivity Range 0 to 200 mS
Resolution / Accuracy 0.01 μS to 0.1 mS / ±1% full scale
Calibration Auto/manual;1 point / range
TDS Range 0 to 200 ppt (TDS factor 1.0)
Resolution / Accuracy 0.01 ppm to 0.1 ppt / ±1 % full scale
Calibration Manual; 1 point / range
Temperature Range -10.0 to 110.0 ºC (14.0 to 230.0 ºF)
Resolution / Accuracy 0.1 0C/ 0F / ± 0.5 ºC; ± 0.9 ºF
Calibration Offset in 0.1 increments, ±10 ºC or ±18 ºF
Display Custom LCD with backlight
Memory Stores up to 500 data sets
Output USB , RS232 output with optional data cable
Real-Time Clock Time and date stamp on Calibration and stored data
Power Two AA batteries (included), >100 hours (without backlight);
AC adapter (optional)
Standard supply Meter, double-junction pH electrode, conductivity cell (K = 1.0), Grip-Clip™, two AA batteries, hard carry case, USB cable, calibration solutions
Dimensions
(L x W x H)
Meter 77/8” x 31/8” x 21/4” (20 x 8.3 x 5.7 cm)
Boxed 91/8” x 91/8” x 23/4” (23 x 23 x 7 cm)
Weight Meter 1 lb (0.5 kg)
Boxed 1.4 lb (0.7 kg)

Ordering Information

Order No. Item
ECPCWP45004K PC 450 Meter with separate pH & cond probes Kit with double-junction pH electrode ECFC7252201B, conductivity cell CONSEN91W (K = 1.0) , Grip-Clip™, two AA batteries, hard carry case, USB cable and calibration solutions
60X426401 Optional 100 / 240 VAC SMPS power adapter, 9 V, 6 W
ECFC7960201B General purpose plastic-body double junction gel-filled ORP Electrode, 12x90mm, BNC connector, 1m cable length
ECFC7960101B General purpose plastic-body single junction gel-filled ORP Electrode, 12x90mm, BNC connector, 1m cable length

pH Calibration Solutions

Order No. Item
ECBU1BT pH 1.68 buffer, 480 ml
ECBU4BTC pH 4.01 buffer, 480 ml
ECBU7BTC pH 7.00 buffer, 480 ml
ECBU10BTC pH 10.01 buffer, 480 ml
ECBU12BT pH 12.45 buffer, 480 ml

ORP Calibration Solutions

Order No. Item
ECORPQUIN Quinhydrone 263 mV, 480 ml
ECORPQUIN086 Quinhydrone 86 mV, 480 ml

Conductivity Calibration Solutions

Order No. Item
ECCON100BT 100 μS KCL, 480 ml
ECCON500BT 500 μS KCL, 480 ml
ECCON1413BT  1413 μS KCL, 480 ml
ECCON2764BT  2764 μS KCL, 480 ml
ECCON5000BT 5.0 mS KCL, 480 ml
ECCON1288BT 12.88 mS KCL, 480 ml
ECCON1118BT 111.8 mS KCL, 480 ml

TDS Calibration Solutions

Order No. Item
EC44250BT 50 ppm, 480 ml
EC442300BT 300 ppm, 480 ml
EC4421000BT 1000 ppm, 480 ml
EC4423000BT 3000 ppm, 480 ml