pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample

The selection of a suitable electrode depends on the types of application

Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable

Single Junction

Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers

Double Junction

Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes

Sealed(gel-filled)

No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required

Refillable

Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low

Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses

pH Electrodes (General Glass) & (General Plastic)

 

SKU: f001-1-1-2-1-3-1-1-1-2-1-2-1-1-1-1-1-1-2-1-2-2-2-1-1-2-1-1-1-1-1-1-1-1-1 Categories: ,

Description

AN-UV-2500N spectrophotometer, equipped with 6 inches LCD display, is an ideal and advanced analytical instrument for laboratory to realize wavelength scanning, Kinetics test, multi wavelength functions. All functions can be operated on spectrophotometer and can be read directly on the display

6 Inches LCD Display
  • AN-UV-2500N series has a 6 inches LCD display to show results and curves directly on the screen
    Powerful Functions
  • Multi functions like spectrum scanning, standard curve, kinetics, multi wavelength, DNA/Protein testing can be operated directly on the spectrophotometer and all corresponding curves and data can be displayed directly
    Perfect Calibration System
  • All baseline, wavelength, dark current can be calibrated automatically to keep good running conditions
    Lead Screw Structure
  • The instrument has a lead screw structure to make the rotation angle of the grating perfectly, so that the instrument wavelength accuracy and wavelength resolution can be greatly improved
    8mm Thick Optical Base
  • AN-UV-2500N uses a rigid 8mm die-cast aluminum base as its optical mount to ensure instrument stability and reliability
    6, Data Output
  • AN-UV-2500N is equipped with USB port to connect with a PC to display spectrum scanning, kinetics and multi wavelength testing results on the screen. The software is optional

Product Description

Specifications AN-UV-2500N
 Wavelength Range 190-1100 nm
 Bandwidth 1.8 nm
 Wavelength Accuracy ±0.5 nm
 Wavelength Repeatability 0.3 nm
 Photometric Accuracy ±0.3% T
 Photometric Repeatability 0.2% T
 Photometric Range -0.3-3A,0-200%T,0-9999 C
 Stability ± 0.002 A/h @ 500 nm
 Baseline Flatness ± 0.002 A
 Noise  ± 0.001 A
 Stray Light ≤0.05% T @ 220 nm,360 nm
 Data Output Port USB
 Printer Port Parallel Port
 Display 320*240 Dots LCD
 Lamps D2  Lamp & W  Lamp
 Detector Silicon Photodiode
 Power Requirements AC   220V/50Hz
 Dimension 460*380*180 mm
 Weight 20 kg

Ordering Information

Order No. Item

AN-UV-2500N

Single beam UV/Vis spectrophotometer with 1 cm glass cuvette ( 4 pcs), 1 cm quartz cuvette(2 pcs) , power cord AC 220V/50Hz and dust cover with software