ION 6+

Product Description

pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample

The selection of a suitable electrode depends on the types of application

Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable

Single Junction

Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers

Double Junction

Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes

Sealed(gel-filled)

No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required

Refillable

Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low

Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses

pH Electrodes (General Glass) & (General Plastic)

 

SKU: f001-1-1-2-1-3-1-1-1-1 Categories: ,

Description

  • Sturdy rubber boot protects the meter against accidental knocks
  •  3-point push button calibration
  •  Direct ion concentration measurement with ion selective electrodes. Also measures pH
  •  Choice of US, NIST and Pure low ionic pH buffer standards
  •  Auto buffer recognition minimizes calibration error
  •  Self-diagnostic messages

Product Description

Product Specification Description
Ion Concentration Range 0.01 to 0.99 ppm / 1.0 to 199.9 ppm / 200 to 1999 ppm
Resolution / Accuracy 0.01 ppm (0.01 to 0.99 ppm) ; 0.1 ppm (1.0 to 199.9 ppm) ; 1 ppm (200 to 1999 ppm) ; ± 1% of Full scale
mV Range ±  500 mV
Resolution / Accuracy 1 mV (±199.9); 1 mV (beyond) /± 0.2 mV (±199.9); ± 2 mV (beyond)
pH Range 0.00 to 14.00 pH
Resolution/Accuracy 0.01 pH  /   ±  0.01 pH
Temperature Range 0.0 to 100.0 °C
Resolution/ Accuracy 0.1 °C /  ±  0.5 °C
No. of Ion Calibration Points Up to 3 points
No. of pH Calibration Points Up to 3 points
No. of pH Buffer Options USA : pH 4.01,7.00, 10.01 ,NIST: pH 4.01, 6.86, 9.18
PWB : pH 4,10, 6.97
pH Electrode Included ECFC7252101B, single junction pH electrode
Ion Electrodes To be ordered separately
Temperature Probe Included
Carrying case Yes
Power Requirement 4 1.5 V ‘AAA’ batteries, battery life > 200 hrs.
Mains operation not available
Dimensions
(L x W x H)
Meter 15.7 x 8.5 x 4.2 cm
Boxed 24 x 17 x 8 cm
Weight Meter 255 g
Boxed 555 g

Ordering Information

Order No. Item
ECION602PLUSK Ion 6+ pH/Ion/ORP Meter with single junction pH electrode ECFC7252101B, ATC probe & pH carrying kit set

pH Calibration Solutions

ECBU4BTC pH 4.01 buffer, 480 ml
ECBU7BTC pH 7.00 buffer, 480 ml
ECBU10BTC pH 10.01 buffer, 480 ml
EC-BU-6.86BT pH 6.86 buffer, 480 ml
EC-BU-9.18BT pH 9.18 buffer, 480 ml

Download Brochure

• Eco Plus Series Family Brochure