PD 650

Product Description

pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample

The selection of a suitable electrode depends on the types of application

Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable

Single Junction

Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers

Double Junction

Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes

Sealed(gel-filled)

No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required

Refillable

Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low

Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses

pH Electrodes (General Glass) & (General Plastic)

 

SKU: f001-1-1-2-1-3-1-1-1-2-1-2-1-1-1-1-1-1-2-1-2-2-2-1-1-2-2-1 Categories: ,

Description

  • Simultaneously measures and displays three parameters view individual parameter details at the press of a button
  • Measures accuracy of up to ±0.002 pH and resolution of up to 3-decimal points
  • Ion measurement capabilities with 3 digit resolution and 8 calibration points
  • Measures 90.00 mg/L in DO concentration and 600 % in DO saturation – one of the widest Dissolved Oxygen measurement ranges offered in the handheld market
  • Rugged and waterproof for applications in harsh environments
  • Step-by-step prompts that guide users through set-up, calibration and trouble-shooting
  • High/low alarm limits
  • Meter automatically logs up to 500 readings with time and date in GLP-compliant format
  • RS 232 through LED, USB / IrDA Interface Adapter
  • Non-volatile memory protects information and meter settings, even when batteries run out
  • Password protection to prevent tampering

Product Description

Product Specification Description
pH Range -2.000 to 20.000 pH
Resolution / Accuracy 0.1 / 0.01 pH / 0.001 pH ; ±  0.002 pH
Calibration Points 1 (Offset) to 6 points
Auto-Buffer Recognition Yes
Slope & Offset Display Yes
Ion Range 0.001 to 19900 ppm
Resolution / Accuracy 2 or 3 digits / 0.5 % Full scale (monovalent)
1 % Full scale (divalent)
Calibration Points Up to 8
mV Range ± 2000.0 mV
Resolution / Accuracy 0.1 mV / ±0.2 mV
Temperature Range -10.0 to 110.0 0C / 14.0 to 230.0 0F (Meter)
Resolution / Accuracy 0.1 0C/0.1 0F / ±0.5 0C/ ±0.9 0F
Conductivity Range 0 to 2.000 ,2.000 to 300.0  μS/cm;
300.0 μS/cm to 4.000 mS/cm
4.000 to 40.00 mS/cm,40.00 to 500.0 mS/cm
Resolution / Accuracy 0.01, 0.1 μS/cm ; 0.001,0.01,0.1 mS/cm / ±1 % Full scale + 1 LSD
Calibration Points 4 (1 per range) auto,
5 (1 per range) manual
Cell Constant (K) 0.010 to 10.000
Temperature Coefficient Linear & pure
Normalisation Temperature 15 to 30 0C
Auto-Ranging Yes
TDS Range 0 to 2.000,2.000 to 300.0 ppm;
300.0 ppm to 4.000 ppt
4.000 to 40.00 ppt,40.00 to 500.0 ppt
Resolution / Accuracy 0.01, 0.1 ppm ; 0.001,0.01,0.1 ppt / ±1 % Full scale + 1 LSD
Calibration Points 5 points
Conversion Factor 0.4000 to 1.000
Auto-Ranging Yes
Salinity Range 0 to 0.770,0.770 to 143.3 ppm;
143.3 ppm to 2.138 ppt
2.138 to 23.64 ppt,23.64 to 80.00 ppt
Resolution / Accuracy 0.01, 0.1 ppm ; 0.001,0.01 ppt / ±1 % Full scale + 1 LSD
Resistivity Range 2.000 Ω to 25.00 Ω, 25.00 Ω to 250.0 Ω, 250.0 Ω to 3.333 kΩ, 3.333 kΩ to 500.0 kΩ, 500.0 kΩ to 20.00 MΩ
Resolution / Accuracy 0.01Ω , 0.1 Ω ,0.001 Ω,0.1 Ω,0.01MΩ / ±1 % Full scale + 1 LSD
Dissolved Oxygen Range 0.00 to 90.00 mg/L or ppm
Resolution / Accuracy 0.01 mg/L ; 0.01 ppm / ± 0.2 mg/L
Calibration Points 1
Probe Galvanic with integral temperature sensor
% Saturation of Oxygen Range 0 to 600.0 %
Resolution / Accuracy 0.1 % / ± 2.0 %
Salinity Correction Range 0.0 to 50.0 ppt
Resolution 0.1 ppt
Method Automatic /manual
Barometric Pressure Correction 450 to 825 mmHg or 59.9 to 109.9 kP
Resolution 0.1 mmHg or 0.1 kPa
Method Automatic correction with built-in sensor
Operating Temperature 0 to 50 0C
ATC Yes
Hold Function Yes
Ready Function Yes
Memory 500 Data sets
Inputs DC Phono sockets, BNC, 8 Pin connector, 6 pin connector
Communication Output USB / IrDA Interface Adapter (for 600 Series Meter to PC Communication) to be ordered separately
Power 4 x 1.5 V ‘AA’ Alkaline batteries or 9 V DC Adapter, 500 mA
Standard supply pH electrode ECFC7252203B, 4 cell Conductivity probe ECCONSEN9203J K= 0.530 , Dissolved Oxygen probe ECDOHANDYNEW, DAS software, power adapter & carrying kit set (all probes have  3 m cable length) ORP electrode and Ion Selective Electrodes to be ordered separately
Dimensions
(L x W x H)
Meter 18.3 x 9.5 x 5.7
Boxed 23.5 x 16 x 15.5 cm
Weight Meter 460 g
Boxed 800 g

Ordering Information

Order No. Item
ECPCDWP65044K Waterproof Cyber Scan PCD 650 pH/ mV/ Ion/ Conductivity/ TDS/ Resistivity/ Salinity/ Dissolved Oxygen Meter with double junction pH electrode ECFC7252203B, 4-cell Conductivity probe ECCONSEN9203J, Dissolved Oxygen probe ECDOHANDYNEW, DAS software, multi-probe holder, power adapter & carrying kit set (all probes comes with 3m long cable).
01X447602 USB / IrDA Interface Adapter  (for 600 Series Meter to PC Communication)
ECFC7960201B General purpose plastic-body double junction gel-filled ORP Electrode, 12x90mm, BNC connector, 1m cable length
ECFC7960101B General purpose plastic-body single junction gel-filled ORP Electrode, 12x90mm, BNC connector, 1m cable length

ORP Calibration Solutions

Order No. Item
ECORPQUIN Quinhydrone 263 mV
ECORPQUIN086 Quinhydrone 86 mV

pH Calibration Solutions

Order No. Item
ECBU1BT pH 1.68 buffer, 480 ml
ECBU4BTC pH 4.01 buffer, 480 ml
ECBU7BTC pH 7.00 buffer, 480 ml
ECBU10BTC pH 10.01 buffer, 480 ml
ECBU12BT pH 12.45 buffer, 480 ml
EC-BU-6.86BT pH 6.86 buffer, 480 ml
EC-BU-9.18BT pH 9.18 buffer, 480 ml

Conductivity Calibration Solutions

Order No. Item
ECCON100BT 100 μS KCL, 480 ml
ECCON500BT 500 μS KCL, 480 ml
ECCON1413BT 1413 μS KCL, 480 ml
ECCON2764BT 2764 μS KCL, 480 ml
ECCON5000BT 5.0 mS KCL, 480 ml
ECCON1288BT 12.88 mS KCL, 480 ml
ECCON1118BT 111.8 mS KCL, 480 ml

TDS Calibration Solutions

Order No. Item
EC44250BT 50 ppm, 480 ml
EC442300BT 300 ppm, 480 ml
EC4421000BT 1000 ppm, 480 ml
EC4423000BT 3000 ppm, 480 ml

Salinity Calibration Solutions

Order No. Item
ECNACL5PPT 5 ppt Nacl, 480 ml
ECNACL25PPT 25 ppt Nacl, 480 ml