Salt 6+

Product Description

pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample

The selection of a suitable electrode depends on the types of application

Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable

Single Junction

Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers

Double Junction

Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes

Sealed(gel-filled)

No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required

Refillable

Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low

Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses

pH Electrodes (General Glass) & (General Plastic)

 

Description

  • Direct readout meter for quick measurement of salinity
  • Automatic Temperature Compensation
  • User customization of Normalisation Temperature and Temperature Co-efficient permits optimum performance
  • Single point push button calibration
  • Manual calibration allows non-standards solutions to conform to individual‘s requirements
  • Unit of measurement can be toggled between % and ppt at the touch of a button to suit individual requirement
  • Freeze and auto-power off function
  • Rubber boot protects the meter from damage due to accidental knocks

Product Description

Product Specification Description
Salinity Range( ppt) 1.0 to 50 ppt
Resolution/Accuracy 0.1 ppt ; ±1% full scale
Salinity Range (%) 0.1% to 5.00%
Resolution/Accuracy 0.01% ; ±1 % full scale
Temp. Range 0 to 80 ºC
Resolution/Accuracy 0.1 0C ; ± 0.5 0C
Temperature Coefficient 0.0 to 3.0 %
Normalisation Temp. 15 ºC fixed
Calibration Points 1 Point
Electrode Included, ultem body ECCONSEN91B, cell constant 1.0, 1 m cable
Carrying Case Yes
Power Requirement 4 x 1.5 V ‘AAA’ batteries; battery life >100 hours
Dimensions(L x W x H) Meter 15.7 x 8.5 x 4.2 cm
Boxed 24 x 17 x 8 cm
Weight Meter 255 g
Boxed 555 g

Ordering Information

Order No. Item
ECSALT603PLUSK Salt 6+ Salinity Meter with electrode ECCONSEN91B & Salinity carrying kit set

Salinity Calibration Solutions

Order No. Item
ECNACL5PPT  5 ppt Nacl, 480 ml
ECNACL25PPT 25 ppt Nacl, 480 ml
ECNACL45PPT 45 ppt Nacl, 480 ml