TDS 6+

Product Description

pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample

The selection of a suitable electrode depends on the types of application

Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable

Single Junction

Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers

Double Junction

Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes

Sealed(gel-filled)

No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required

Refillable

Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low

Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses

pH Electrodes (General Glass) & (General Plastic)

 

Description

  • Sturdy rubber boot protects the meter against accidental knocks
  • Multi-point push button calibration
  • User selectable TDS factor (0.4 to 1.0)
  • Auto-ranging capability, quickly detects appropriate measurement range
  • User customization of functions like normalization temperature, temperature co-efficient, automatic or manual calibration etc
  • Auto Temperature Compensation
  • Hold and auto-off functions
  • Self diagnostic messages

Product Description

Product Specification Description
TDS Range 0- 9.99, 10-99.9, 100-999 ppm; 1.0-9.99 , 10- 99.9 ppt Max.199.9 ppt depending on factor setting
Resolution/Accuracy 0.01, 0.1 ,1 ppm; 0.01,0 .1 ppt (0.05% of full scale ) ;
± 1% of full scale
Calibration Points 5; Maximum 1 per range
Temp. Range 0 to 100 0C ( 32 to 212 0F)
Resolution/Accuracy  0.1 0C / 0.1 0F  ; ± 0.5 0C / ± 0.9 0F
Temp.Compensation ATC / MTC (0 to 80 0C )
TDS Factor 0.4 to 1.0
Temp. Coefficient 0.0- 3.0%
Normalisation Temp. 20 or 25 0C
Electrode Included, ultem body ECCONSEN91B, cell constant 1.0, 1m cable
Carrying case Yes
Power Requirements 4 x 1.5 V ‘AAA’ batteries, battery life > 100 hrs.
Mains operation not available
Dimensions(L x W x H) Meter 15.7 x 8.5 x 4.2 cm
Boxed 24 x 17 x 8 cm
Weight Meter 255 g
Boxed 555 g

Ordering Information

Order No. Item
ECTDS603PLUSK TDS 6+ TDS Meter with electrode ECCONSEN91B & TDS carrying kit set

TDS Calibration Solutions

Order No. Item
EC44250BT 50 ppm, 480 ml
EC442300BT 300 ppm, 480 ml
EC4421000BT 1000 ppm, 480 ml
EC4423000BT 3000 ppm, 480 ml