Turbidity Meter TN100

Turbidity Meter TN100

Product Description

pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample

The selection of a suitable electrode depends on the types of application

Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable

Single Junction

Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers

Double Junction

Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes

Sealed(gel-filled)

No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required

Refillable

Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low

Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses

pH Electrodes (General Glass) & (General Plastic)

 

SKU: f001-1-1-2-1-3-1-1-1-2-1-2-1-1-1-1-1-1-2-1-2-2-2-1-1-2-1-1-1-1-1-1-1-1-1-1-1-1-2-2-1-1-1-3-1-1-1-2-1-1 Categories: ,

Description

  • IP67 Waterproof and Dustproof – Cleaning and maintenance is simple. It floats!
  • Automatic Push-Button Calibration (4 points) for unprecedented full-range accuracy
  • Sturdy Carrying Kit complete with 4 calibration standards, 3 empty cuvettes, batteries and comprehensive instruction manual
  • Power-saving feature allows over 1200 measurements to be taken with a single set of AAA alkaline batteries
  • Large Easy-to-Read Display for optimum viewing of measurements with user-friendly message codes for easy troubleshooting
  • Meets Performance Criteria as specified by ISO 7027 (DIN EN 27027) method

Product Description

Product Specification Description
Measuring Parameter Turbidity
Highlights Light weight, IP67 waterproof meter
Principle Nephelometric non-ratio/ ISO 7027 and DIN 27027 compliant
Range 0 to 2000 NTU
Automatic Range Selection 0.01 to 19.99 NTU; 20.0 to 99.9 NTU; 100 to 2000 NTU
Resolution 0.01 NTU (0 to 19.99 NTU); 0.1 NTU (20.0 to 99.9 NTU); 1 NTU (100 to 2000 NTU)
Accuracy < 0.1 NTU for 0.02 NTU standard; ±2 % of reading ±1 digit for 0 to 500 NTU; ±3 % of reading ±1 digit for 501 to 2000 NTU
Repeatability < ±1 % of reading
Calibration Up to 4 points
Calibration Standards 0.02 NTU, 20.0 NTU, 100 NTU, 800 NTU
Response Time < 6 secs for full step change
Sample Volume 10 mL
Light Source Infrared-emitting diode (850 nm wavelength)
Operating Temperature Range 0 to 50 0C
Sample Temperature Range 0 to 50 0C
Enclosure Type & Rating ABS plastic & IP67 rated
Power 4 x 1.5 V ‘AAA’ alkaline batteries(> 1200 readings)
Certifications/Compliance FCC part 15 class A, TUV 3 in 1, CE, RoHS
Dimensions(L x W x H) Meter 15.5 x 6.8 x 4.6 cm Boxed 35.5 x 16.5 x 10.5 cm
Weight Meter 200 g Boxed 1160 g

Ordering Information

Order No. Item
ECTN100IR Waterproof TN-100 handheld turbidity meter with set of 4 calibration standards (800, 100, 20.0 & 0.02 NTU), sample vials, lint-free cloth, silicone oil, batteries, and hard carrying case
TN100CALKIT Calibration kit set – standard solutions (0.02, 20.0, 100, 800 NTU) 60 mL each and (4) vials
ECTN100CUVKIT Pack of 3 sample cuvettes (vials)
ECTN100CUVKIT4 Pack of 4 cuvettes (vials)
ECSILICONEOIL Silicone Oil (10 ml) for TN100IR
01X344318 10.0 NTU Calibration Standard (125 mL)