• Up to 6-point calibration with auto-buffer recognition
    • Quick, easy electrode diagnosis with multiple pH slopes and offset display
    • Non-volatile memory holds up to 500 data points – time and date-stamped for GLP compliance
    • Calibrate with up to 5 custom pH buffers – use any pH values that are ≥1.0 pH unit apart
    • Cal-due alarm – no more out-dated calibrations
    • Auto-logging function for convenient continuous monitoring
    • Password protection for setup and calibration

    Product Description

    Product Specification Description
    pH Range -2.000  to  20.000 pH
    Resolution/Accuracy 0.1 / 0.01/ 0.001 pH (selectable) /± 0.002 pH + 1 LSD
    mV Range ± 2000 mV
    Resolution/Accuracy 0.1 mV / ± 0.2 mV
    Temp.Range 0.0 to 100.0 ºC / 32.0 to 212.0 ºF
    Resolution/Accuracy 0.1 ºC (0.1 ºF) / ± 0.3 ºC (0.5 ºF)
    ATC Yes (0 to 100 ºC)
    No.of pH Calibration Points 6 points
    Buffer Sets USA, NIST, DIN, User 1, User 2, Custom
    Built-in Electrode Arm Yes
    GLP Yes
    Memory 500 data sets
    Output RS 232C (via cable)
    Electrode Included ECFG7370101B, Single junction glass pH electrode, ATC Probe (PH5TEMB01P). ORP electrode to be ordered separately
    Temp.Probe Included
    Power Requirements Included, 100 /240VAC SMPS Power Adapter , 9V,6W
    Dimensions (L x W x H) Meter 17.5 x 15.5 x 6.9 cm Boxed 30.8 x 23.5 x 12.4 cm
    Weight Meter 650 g Boxed 1800 g

    Ordering Information

    Order No. Item
    ECPH270042GS Eutech pH 2700 Meter With pH Electrode (ECFG7370101B), ATC Probe (PH5TEMB01P), Integral Electrode Holder, 100/240 VAC Adapter, RS232 cable, & pH Electrode Refill Solution
    ECFC7960201B General purpose plastic-body double junction gel-filled ORP Electrode, 12x90mm, BNC connector, 1m cable length
    ECFC7960101B General purpose plastic-body single junction gel-filled ORP Electrode, 12x90mm, BNC connector, 1m cable length

    pH Calibration Solutions

    Order No. Item
    ECBU1BT pH 1.68 buffer, 480 ml
    ECBU4BTC pH 4.01 buffer, 480 ml
    ECBU7BTC pH 7.00 buffer, 480 ml
    ECBU10BTC pH 10.01 buffer, 480 ml
    ECBU12BT pH 12.45 buffer, 480 ml
    EC-BU-6.86BT pH 6.86 buffer, 480 ml
    EC-BU-9.18BT pH 9.18 buffer, 480 ml

    ORP Calibration Solutions

    Order No. Item
    ECORPQUIN Quinhydrone 263 mV, 480 ml
    ECORPQUIN086 Quinhydrone 86 mV, 480 ml

    Download Brochure

    • Eutech 2700 Series Family Brochure
  • Product Description

    pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample The selection of a suitable electrode depends on the types of application Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable
    Single Junction
    Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers
    Double Junction
    Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes
    Sealed(gel-filled)
    No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required
    Refillable
    Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses pH Electrodes (General Glass) & (General Plastic)  
  • Product Description

    pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample The selection of a suitable electrode depends on the types of application Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable
    Single Junction
    Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers
    Double Junction
    Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes
    Sealed(gel-filled)
    No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required
    Refillable
    Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses pH Electrodes (General Glass) & (General Plastic)  
  • Product Description

    pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample The selection of a suitable electrode depends on the types of application Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable
    Single Junction
    Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers
    Double Junction
    Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes
    Sealed(gel-filled)
    No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required
    Refillable
    Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses pH Electrodes (General Glass) & (General Plastic)  
  • Product Description

    pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample The selection of a suitable electrode depends on the types of application Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable
    Single Junction
    Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers
    Double Junction
    Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes
    Sealed(gel-filled)
    No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required
    Refillable
    Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses pH Electrodes (General Glass) & (General Plastic)  
  • Product Description

    pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample The selection of a suitable electrode depends on the types of application Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable
    Single Junction
    Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers
    Double Junction
    Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes
    Sealed(gel-filled)
    No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required
    Refillable
    Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses pH Electrodes (General Glass) & (General Plastic)  
  • Product Description

    pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample The selection of a suitable electrode depends on the types of application Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable
    Single Junction
    Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers
    Double Junction
    Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes
    Sealed(gel-filled)
    No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required
    Refillable
    Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses pH Electrodes (General Glass) & (General Plastic)  
  • Product Description

    pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample The selection of a suitable electrode depends on the types of application Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable
    Single Junction
    Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers
    Double Junction
    Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes
    Sealed(gel-filled)
    No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required
    Refillable
    Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses pH Electrodes (General Glass) & (General Plastic)  
  • Product Description

    pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample The selection of a suitable electrode depends on the types of application Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable
    Single Junction
    Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers
    Double Junction
    Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes
    Sealed(gel-filled)
    No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required
    Refillable
    Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses pH Electrodes (General Glass) & (General Plastic)  
  • Product Description

    pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample The selection of a suitable electrode depends on the types of application Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable
    Single Junction
    Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers
    Double Junction
    Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes
    Sealed(gel-filled)
    No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required
    Refillable
    Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses pH Electrodes (General Glass) & (General Plastic)  
  • Product Description

    pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample The selection of a suitable electrode depends on the types of application Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable
    Single Junction
    Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers
    Double Junction
    Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes
    Sealed(gel-filled)
    No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required
    Refillable
    Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses pH Electrodes (General Glass) & (General Plastic)  
  • Product Description

    pH measurement is usually done with the use of a combination electrode. The combination electrode is an electrode system formed by a glass sensing half-cell and an internal reference half-cell. As the reference junction acts as the medium of conductor between the reference electrolyte and the sample to be measured, it must allow free movement of electrons through the junction and into the sample The selection of a suitable electrode depends on the types of application Eutech’s pH /ORP combination electrodes offer the convenience of having the reference and measuring electrodes combined in a single housing. They are offered in a variety of configurations to suit most laboratory and field application needs. Choose from two models: single-or double-junction, sealed (gel-filled) or refillable
    Single Junction
    Ideal for general-purpose applications. Ag+ ions are in contact with junction and this can cause chemical interaction with sulphur. Not suitable for biological samples or tris buffers
    Double Junction
    Prevents interference between the inner fill solution and sample. Electrolyte is free of Ag+ ions. Suitable for use with biological samples. Can be used in place of calomel reference electrodes
    Sealed(gel-filled)
    No refill hole. The reference electrolyte is usually a gel. Replacement of the electrode is necessary when it gets contaminated. Virtually no maintenance is required
    Refillable
    Refill hole allows reference electrolyte replenishment. Can be used many times over. Needs refilling when fill solution is low Choose single junction ORP electrodes for field, clean water, and other general-purpose applications or double junction electrodes for samples containing heavy metals and organics. Platinum sensors are suitable for most general uses pH Electrodes (General Glass) & (General Plastic)